NeoQi equipment is designed for use in indoor conditions and cannot be transported or stored below 0 degrees C. The equipment should be used in a moisture proof room, ventilation is recommended, as in the working regime, the machine creates humidity.
It is recommended to have a general drain on the floor.
Doors must be wide at least 36in.
NeoQi® equipment cannot be used if the air contains FLAMMABLE ANAESTHETIC MIXTURE WITH AIR or WITH OXYGEN OR NITROUS OXIDE.
Maximum allowed load on the vibrating bed is 350lbs.
Electric installations of NeoQi equipment should be carried out by qualified specialists or official representative of NeoQi USA
NeoQi equipment uses 240v & 3 x 2.5 mm² copper crosscut power cord. The power cord should be connected into the moisture proof junction box, which is located on the frame of the equipment.
Maximum power consumption varies depending on the model, ranging from 0.3 up to 4.6 kW.
Protective device current in front of equipment feeder should not be less than 20A.
The fuse rating used in the main external power supply circuit of stationary installed equipment: circuit breakers 20A.
The installation of warm and cold water inlets and drainage connections of NeoQi USA Harmony equipment should be carried out by qualified specialist or official representative of NeoQi USA
Two water pressure hoses are required – one for hot, another for cold water. Depending on the model, the endings should be either ½ or 3⁄4 threaded. The length of the hoses depend upon the distance from the water inlets.
NeoQi equipment uses Ø 32 or 50 mm connections for the drainage on the floor and the water drainage hose, depending on the model being installed. The length of the drainage hose will depend on the distance from the drainage opening on the ground.
For proper functioning of the equipment, only clean tap water should be used.
In case the tap water contains a high percentage of abrasive substances or rust, it is recommended to install a water filter.
Minimum required water level to operate the hydrotherapy tub should be ~ 1 cm above the back jets.
Before using the equipment and composing the maintenance plan, the general quality and hardness of the water must be assessed.
Before using the equipment and composing the maintenance plan, the general quality and hardness of the water must be assessed.
The warranty will be voided if maintenance is carried out non-systematically and only upon necessity.
Regular maintenance for certain parts is essential for long life and faultless functioning of the equipment. Routine maintenance ensuring safe operation includes cleaning, disinfection and visual control of general condition of the equipment.
Routine maintenance consists of cleaning the water system of the bath and the steam generator of the cocoon.
In case of low quality water (high contents of limestone, iron, abrasive particles etc.), the installation of an appropriate water filter is required. The frequency of cleaning the steam generator is directly dependent on the hardness of water.
Try to avoid scratching the acrylic surface with hard elements (sand, jewelry, accessories, etc.). Do not drop heavy and hard objects on the surface of the device.
Water system cleaning
The cleaning of the water system of the bath (jets, pumps, and pipes) needs to be carried out as required, but no less than after every 50 sessions.
Steam generator cleaning
In order to compose the maintenance plan for the steam generator cleaning, the general quality and hardness of the water must be assessed. We recommend performing a Water test.
SERVICE SCHEDULE
53 ppm CaCO3 Very soft water After every 500 work hours
71 ppm CaCO3 Soft water After every 150 work hours
125 ppm CaCO3 Semi hard water After every 50 work hours
249 ppm CaCO3 Hard water After every 15 work hours
374 ppm CaCO3 Very hard water Install water filter